The Swiss Federal Council, based on the Therapeutic Products Act of 15 December 20001 (TPA), ordains: |
Chapter 1 General Provisions |
Section 1 Scope and Exceptions |
Art. 1 Scope
1This Ordinance applies to:
2 In this Ordinance, the term devices is used to designate the products defined in paragraph 1. 3 This Ordinance also applies to:
|
Art. 2 Exceptions
1 This Ordinance does not apply to:
2 In the cases specified in paragraph 1 letters f–h and j, the part of the combination that is deemed to be a device must satisfy the general safety and performance requirements set out in Article 6.9 8 Inserted by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). 9 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Section 2 Definitions and References to European Legislation |
Art. 3 Medical device and accessories
1 Medical devicesareinstruments, apparatus, appliances, softwares, implants, reagents, materials or other articles:
2Medical devices also include:
3 Accessory for a medical device means an article which, whilst not being itself a medical device, is intended by its manufacturer to be used together with one or more particular medical devices and:
10 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Art. 4 Further definitions
1 In this Ordinance:
2 The definitions set out in Article 2 numbers 3–26, 31, 37, 38, 40–44, 46, 48, 51–53, 57–69 and 71 EU-MDR, taking account of the amendments to the definitions in Article 2 numbers 18–21 EU-MDR, adopted by the European Commission by means of delegated acts13. 11 Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EEC, OJ L117 of 5.5. 2017, p. 1; last amended by Regulation (EU) 202/561, OJ L 130 of 24.4.2020, p. 18. 12 Inserted by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). 13 See Annex 4. |
Art. 5 References to European legislation
1 The equivalent terms specified in Annex 2 and as used in EU-MDR14 and this Ordinance apply. 2 Where this Ordinance makes reference to provisions of EU-MDR that in turn refer to other provisions of EU-MDR or other EU legislative acts, those provisions also apply. The version in the footnote to Article 4 paragraph 1 letter f is authoritative for references to EU-MDR, while the versions of the relevant EU act set out in Annex 3 number 1 apply to references to other EU acts. This provision excludes onward references to the EU acts listed in Annex 3 number 2; here the Swiss provisions listed in the Annex apply. 14 See the footnote to Art. 4 para. 1 let. f. |
Chapter 2 Making available on the Market and Putting into Service |
Section 1 Requirements |
Art. 6 General safety and performance requirements
1 A device may be placed on the market or put into service only if it complies with this Ordinance when duly supplied and properly installed, maintained and used in accordance with its intended purpose. 2 Devices shall meet the general safety and performance requirements set out in Annex I to EU-MDR15, taking account of their intended purpose. 3 Appropriate evidence that the part of the combination that is deemed to be a device under the cases set out in Article 2 letters f–h and j fulfils the device requirements must be presented to the competent authority on demand.16 4 If the device complies with the applicable technical standards designated by the Swiss Agency for Therapeutic Products (Swissmedic) or with common specifications, or relevant sections thereof, or with pharmacopoeial requirements in accordance with the Pharmacopoeia Ordinance of 17 October 200117, it is presumed that the device conforms with those requirements of this Ordinance covered by the applicable designated standards or common specifications, or relevant sections thereof or by the pharmacopoeial requirements.18 5 The presumption in paragraph 4 also applies to compliance with the system or process requirements, which, in accordance with this Ordinance, must be fulfilled by economic operators, including the requirements relating to quality management systems, risk management, post-market surveillance systems, clinical investigations, clinical evaluation or post-market clinical follow-up.19 6 Compliance with the common specifications in paragraph 4 is required unless the manufacturer can duly justify that it has adopted solutions that ensure a level of safety and performance that is at least equivalent thereto. The above is subject to Article 8 paragraph 1.20 15 See the footnote to Art. 4 para. 1 let. f. 16 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). 18 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). 19 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). 20 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Art. 7 Distance sales
1 Devices offered in Switzerland by means of an information society service – specifically an online service – that fulfil the conditions set out in paragraph 4 must comply with this Ordinance.21 1bisDevices offered to users in Switzerland online or via some other form of distance sales are considered to be made available on the market.22 2 Similarly, devices that are not placed on the market but are used in the context of a commercial activity, whether in return for payment or free of charge, for the provision of a diagnostic or therapeutic service offered by means of information society services or by other means of communication must also comply with this Ordinance. 3 Upon request by Swissmedic, any natural or legal person offering a device in accordance with paragraph 1 or providing diagnostic or therapeutic services in accordance with paragraph 2 must provide a copy of the declaration of conformity.23 4 A device is deemed to be supplied via an information society service if that service:
5 ...24 21 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). 22 Inserted by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). 23 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). 24 Repealed by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, with effect from 26 May 2022 (AS 2022 291). |
Art. 8 Specific requirements
1 Devices without an intended medical purpose in accordance with Article 1 paragraph 1 letter b must comply with the common specifications stipulated by Swissmedic. 2 Devices that have both a medical and non-medical intended purpose must fulfil both the requirements for devices with an intended medical purpose and the requirements for devices without an intended medical purpose. 3 Devices that are also machines within the meaning of Article 1 of the Machine Ordinance of 2 April 200825 must satisfy the pertinent general safety and health protection requirements of the Machine Ordinance where these requirements are more specific than those of Chapter II of Annex I to EU-MDR26. |
Art. 9 Devices manufactured and used in healthcare institutions
1 Devices manufactured and used solely within healthcare institutions are deemed to have been put into service. Such devices are subject to the pertinent general safety and performance requirements of Annex I to EU-MDR27 but not to any of the other requirements set out in this Ordinance, provided the requirements of Article 5 paragraph 5 letters a–h EU-MDR are fulfilled. 2 Paragraph 1 does not apply to devices manufactured on an industrial scale. 27 See the footnote to Art. 4 para. 1 let. f. |
Art. 10 Custom-made devices
1 Custom-made devices are subject to the requirements of Annex XIII to EU-MDR28. The devices must be accompanied by the statement referred to in Section 1 of Annex XIII to EU-MDR when they are placed on the market. 2 In addition to the procedure under paragraph 1, manufacturers of class III implantable custom-made devices must also carry out a conformity assessment procedure as specified in Chapter I of Annex IX to EU-MDR. Alternatively, they may opt for a conformity assessment in accordance with Part A of Annex XI to EU-MDR. 3 Manufacturers must draw up, keep up to date and keep available for competent authorities the documentation in accordance with Section 2 of Annex XIII to EU-MDR. 28 See the footnote to Art. 4 para. 1 let. f. |
Art. 11 Systems and procedure packs
1 The requirements of Articles 22 and 29 paragraph 2 EU-MDR apply to the placing on the market of systems and procedure packs29. 2 Any natural or legal person who sterilises systems or procedure packs for placing on the market must apply a conformity assessment procedure for the sterilisation process and involve in that procedure a conformity assessment body that is designated under this Ordinance or recognised under an international agreement (designated body). The modalities for doing so are governed by Article 22 paragraph 3 EU-MDR. 3 Any natural or legal person who places the following systems or procedure packs on the market must fulfil the obligations of a manufacturer under Articles 46–50 and conduct the pertinent conformity assessment procedure under Article 23: Systems or procedure packs that:
29 See the footnote to Art. 4 para. 1 let. f. |
Art. 12 Parts and components
1 Any natural or legal person who makes available on the market an item intended to replace an identical or similar integral part or component of a device that is defective or worn in order to maintain or restore the function of the device without changing its performance or safety characteristics or its intended purpose shall ensure that the item does not adversely affect the safety and performance of the device. Supporting evidence must be kept available for the competent authority. 2 An item that is intended specifically to replace a part or component of a device and that significantly changes the performance or safety characteristics or the intended purpose of the device shall be considered to be a device and shall meet the requirements laid down in this Ordinance. |
Art. 13 Conformity marking and identification number
1 Devices placed on the market in Switzerland or made available on the Swiss market must bear a conformity marking in accordance with Annex 5. The conformity marking presented in Annex V to EU-MDR30 is also a permissible conformity marking. 2 The following must not bear a conformity marking:
3 Where devices’ conformity has to be assessed by a designated body, the identification number of this body must be affixed to the conformity marking. 30 See the footnote to Art. 4 para. 1 let. f. 31 SR 812.213.3 32 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Art. 14 Affixing conformity markings and identification numbers
1 The conformity marking and, where necessary, the associated identification number shall be affixed to the device itself or its sterile packaging. 2 Where this is not possible or practicable owing to the nature of the device, the conformity marking and, where necessary, the associated identification number must be displayed on the packaging. 3 The conformity marking shall also appear in the instructions for use and on the sales packaging. 4 The requirements of Article 20 paragraphs 3–6 EU-MDR33 and the general principles of Article 30 Regulation (EC) No. 765/200834 must also be observed when affixing the conformity marking. 33 See the footnote to Art. 4 para. 1 let. f. 34 Regulation (EC) No. 765/2008 of the European Parliament and of the Council of 9 July 2008 setting out the requirements for accreditation and market surveillance relating to the marketing of products and repealing Regulation (EEC) No 339/93, last amended by OJ L 218 of 13.8.2008, p. 30. |
Section 2 Classification, Product Information und Product Identification 35
35 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Art. 15 Classification
1 Devices shall be divided into classes I, IIa, IIb and III, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-MDR36. 2 ...37 36 See the footnote to Art. 4 para. 1 let. f. 37 Repealed by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, with effect from 26 May 2022 (AS 2022 291). |
Art. 16 Product information
1 Product information comprises the labelling and instructions for use. It is governed by Chapter III of Annex I to EU-MDR38. 2 It must be written in all three official languages of Switzerland. Symbols established by means of technical standards may be used to replace written statements. 3 The product information may be provided in fewer than the three official languages of Switzerland or in English, provided that:
4 If requested, additional information must be provided to users in one of the official languages of Switzerland. 5 If a product cannot be, or cannot yet be, placed on the market as a medical device but may be confused with such a device, the claims relating to the said product must indicate clearly and legibly that it is not a medical device and is not suitable for medical purposes. 6 Devices intended solely for demonstration and presentation purposes must be specifically labelled as such. The information must be clearly visible and comprehensible. 7 Misleading or contradictory information on a device’s intended purpose, safety and performance is forbidden. 38 See the footnote to Art. 4 para. 1 let. f. 39 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). 40 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Art. 17 Unique device identification
1 Any manufacturer or any natural or legal person who assembles systems and procedure packs in accordance with Article 22 paragraphs 1 and 3 EU-MDR41 shall assign to the device, system or procedure pack, with the exception of custom-made devices, and all higher levels of packaging a unique device identifier (UDI42) prior to placing it on the market.43 2 It must state the UDI on the labelling of the device, system or procedure pack and all higher levels of packaging. Shipping containers are not considered as a higher level of packaging.44 3 It shall maintain a list of all the UDIs it has assigned. This list is part of the technical documentation specified in Annex II to EU-MDR. It must be kept up-to-date at all times.45 4 The obligations and modalities associated with device identification and registration are governed by Articles 27 and 29 and Annex VI to EU-MDR, taking account of the amendments to this Annex adopted by the European Commission by means of delegated acts46. 5 ...47 41 See the footnote to Art. 4 para. 1 let. f. 42 Stands for «Unique Device Identification» 43 Amended by No I of the O of 19 May 2021, in force since 26 May 2021 (AS 2021 281). 44 Amended by No I of the O of 19 May 2021, in force since 26 May 2021 (AS 2021 281). 45 Amended by No I of the O of 19 May 2021, in force since 26 May 2021 (AS 2021 281). 46 See Annex 4. 47 Enters into force at a later date (Art. 110 para. 2). |
Section 3 Reporting Obligations and Information |
Art. 18 Notification of devices manufactured in healthcare institutions 48
1 Healthcare institutions that manufacture and use devices as specified in Article 9 shall provide the following information to Swissmedic prior to putting the devices into service:
2 Any other relevant information about these devices must be submitted to Swissmedic upon request.49 3 Changes to the information required in paragraph 1 must be reported to Swissmedic within 30 days. 4 Depending on the risk inherent to a device and its use, Swissmedic may exempt devices manufactured and used in accordance with Article 9 from the reporting obligation. 48 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). 49 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Art. 19 Reporting obligation for natural and legal persons who make custom-made devices available on the market
1 Any natural or legal person who makes custom-made devices available on the Swiss market must provide the following information to Swissmedic before making the devices available:
2 Changes to this information must be reported to Swissmedic within 30 days of the changes taking effect. 3 Depending on the risk inherent to a device and its use, Swissmedic may exempt custom-made devices from the reporting obligation under paragraph 1. 50 See Annex 4. |
Art. 20 Information on implantable devices
1 For implantable products, the manufacturer must provide, in addition to the product information required under Article 16, the information required under Article 18 paragraph 1 EU-MDR51, including the implant card. The exemptions specified under Article 18 paragraph 3 EU-MDR apply, taking account of the amendments adopted by the European Commission by means of delegated acts52. 2 The implant card must drawn up in all three official languages of Switzerland. 3 Healthcare institutions must enter the details of the implant recipient in the implant card and give the card to the recipient. They provide the essential information needed by the recipient in a quickly accessible mean. |
Chapter 3 Conformity Assessment, Certificate and Declaration |
Section 1 Conformity assessment |
Art. 21 Principle
1 Any natural or legal person who is domiciled in Switzerland and makes devices available on the market in Switzerland or in a contracting state must, upon request, provide the authorities that are responsible for controls in the field of market surveillance with the declaration of conformity. 2 A manufacturer who places a device on the market in Switzerland or in a contracting state and who is domiciled in Switzerland must carry out an assessment of the device’s conformity with the relevant conformity assessment procedures prior to placing it on the market. The manufacturer and the importer must be able to prove that such a conformity assessment has been carried out and that the device is conforming.53 3A manufacturer who puts into service in Switzerland or in a contracting state a device that is not placed on the market and who is domiciled in Switzerland must carry out an assessment of the device’s conformity with the relevant conformity assessment procedures before it is put into service. The manufacturer must be able to prove that such a conformity assessment has been carried out and that the device is conforming.54 4 The demonstration of compliance with the general safety and performance requirements must also include a performance evaluation in accordance with Article 61 EU-MDR55.56 53 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). 54 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). 55 See the footnote to Art. 4 para. 1 let. f. 56 Inserted by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Art. 22 Exemptions
1 On a duly justified request, Swissmedic may authorise the placing on the market and putting into service of a specific device the use of which is in the interests of public health or patient safety or health even though:
2 Individual devices that have not undergone the relevant conformity assessment procedure may be placed on the market and used without authorisation from Swissmedic provided:
3 For devices placed on the market exclusively within the armed forces or within the framework of their specific tasks, the Federal Department of Home Affairs (FDHA) may, in agreement with the Federal Department of Defence, Civil Protection and Sport, grant exemptions. 57 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Art. 24 Involvement of a designated body
1 When a designated body is involved, all the information necessary for the conformity assessment must be made available to it. 2 Manufacturers must not simultaneously apply to more than one designated body in Switzerland or a contracting state to carry out a conformity assessment procedure for the same product. 3 Any natural or legal person who applies to a designated body must inform that body whether an application to a different designated body in Switzerland or a contracting state has been withdrawn before a decision was issued or rejected by a different designated body in Switzerland or a contracting state. 4 If a manufacturer withdraws its application to have a conformity assessment procedure carried out before the designated body has issued its decision on the assessment, the designated body in question shall notify Swissmedic and the other designated bodies. 5 ...60 6 Where a manufacturer voluntarily changes the designated body, it must comply with the requirements of Article 58 EU-MDR61. 60 Repealed by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, with effect from 26 May 2022 (AS 2022 291). 61 See the footnote to Art. 4 para. 1 let. f. |
Section 2 Certificate of Conformity |
Art. 25 Issuing and content
1 The designated bodies issue certificates of conformity in accordance with Annexes IX–XI to EU-MDR62 (Certificates). 2 The certificates must be issued in one of the three official languages of Switzerland or in English. 3 They must, as a minimum, include the information required in Annex XII to EU-MDR, taking account of the amendments to this Annex adopted by the European Commission by means of delegated acts63. 4 Certificates issued by bodies designated under EU law and domiciled in a state of the EU or European Economic Area (EEA) but not recognised by an international agreement are deemed equivalent to certificates issued by Swiss bodies if it can be credibly demonstrated that:
62 See the footnote to Art. 4 para. 1 let. f. 63 See Annex 4. 64 Inserted by No I of the O of 19 May 2021, in force since 26 May 2021 (AS 2021 281). |
Art. 26 Validity
1 Certificates are valid for a maximum of five years. The expiry date must be indicated on the certificate. 2 At the manufacturer's request, the validity of the certificate may be extended by a maximum of five years following a re-assessment carried out in accordance with the relevant conformity assessment procedure. Certificates may be extended more than once. 3 Any supplement to a certificate is valid for the same period as the certificate to which it belongs. |
Art. 27 Suspension, restriction and revocation
1 If a designated body finds that a manufacturer no longer fulfils the requirements of this Ordinance, it shall impose on that manufacturer a suitable deadline for restoring compliance. 2 If this deadline passes without the manufacturer taking suitable corrective action, the designated body shall suspend, revoke or restrict the certificate in question. 3 A certificate that has been amended, suspended or revoked by a designated body must no longer be used in its original form. |
Art. 28 Documentation requirements
1 The designated body shall provide Swissmedic and the other designated bodies with:65
2 It shall also provide Swissmedic with information on whether or not a conformity assessment procedure should be applied in accordance with Article 54 paragraph 1 EU-MDR66. Notifications of certificates for devices that have undergone a procedure of this type must include the documents specified in Article 55 paragraph 1 EU-MDR.67 65 Amended by No I of the O of 19 May 2021, in force since 26 May 2021 (AS 2021 281). 66 See the footnote to Art. 4 para. 1 let. f. 67 Amended by No I of the O of 19 May 2021, in force since 26 May 2021 (AS 2021 281). |
Section 3 Declaration of Conformity |
Art. 29
1 If the applicable conformity assessment procedure has demonstrated that the requirements of this Ordinance have been fulfilled, the manufacturer of devices that are not custom-made or investigational issues a declaration of conformity. The manufacturer shall continuously update this declaration. 2 The declaration of conformity shall include the information required in Annex IV to EU-MDR68, taking account of the amendments to this Annex adopted by the European Commission by means of delegated acts69. It must be written in one of the three official languages of Switzerland or English or translated into one of these languages. 3 If a device also requires a manufacturer's declaration of conformity for aspects not covered by this Ordinance but nevertheless required by other legislation in order to demonstrate compliance with that legislation, a single declaration of conformity shall be drawn up. 4 By drawing up the declaration of conformity, the manufacturer assumes responsibility for compliance with the requirements of this Ordinance and all other legislation applicable to the device. |
Chapter 4 Requirements applicable to Tissues and Cells of Human Origin that are incorporated into or used in the Production of Devices |
Art. 30 Establishment licence
1 An establishment licence issued by Swissmedic is required for any natural or legal person who:
2 A licence shall be issued if:
3 Swissmedic shall verify that the conditions for issuing an establishment licence are fulfilled in the course of an inspection. 4 Articles 39–43 of the Medicinal Products Licensing Ordinance of 14 November 201870 (MPLO) also apply mutatis mutandis. |
Art. 31 Collection, donation and testing
1 The requirements of Articles 3, 4, 6–15 and 30–33 of the Transplantation Act of 8 October 200471 and of Articles 2–12 of the Transplantation Ordinance of 16 March 200772 apply mutatis mutandis to collection, donation and testing. 2 Holders of an establishment licence issued under Article 30 must verify the suitability of donors. |
Art. 32 Duty to keep records and traceability
1 It must be possible to trace all tissues and cells collected from humans for the purpose of devitalisation and use in devices from the donor to the recipient and vice versa. Article 35, paragraphs 1 and 2, MPLO73 also apply mutatis mutandis. 2 Moreover, the provisions of Articles 34 and 35 of the Transplantation Act of 8 October 200474 apply mutatis mutandis to traceability. |
Chapter 5 Designated Bodies |
Section 1 Designation |
Art. 33 Requirements and application
1 Swissmedic shall only designate conformity assessment bodies that have completed an assessment procedure in accordance with Article 34 and which satisfy the requirements set out in Annex VII to EU MDR75.76 2 Applications for designation must be submitted to Swissmedic. They must in particular include:
3 Swissmedic shall, within thirty days, verify whether the application for designation is complete, and shall request the applicant to provide any missing information. 4 It shall review the application and accompanying documents and issue a preliminary assessment report. 75 See the footnote to Art. 4 para. 1 let. f. 76 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Art. 34 Assessment
1 Swissmedic shall conduct an on-site assessment of the conformity assessment body and, if relevant, of all sub-contractors and subsidiaries. 2 If Swissmedic identifies non-compliances in the course of its assessment, it shall draw up a list of non-compliances for the applicant. Swissmedic shall set the conformity assessment body a deadline by which the latter shall submit to Swissmedic a corrective action plan to remedy the non-compliances and a preventive action plan. 3 The plans shall indicate the root cause of the identified non-compliances and shall include a timeframe for implementation of the actions therein. 4 Swissmedic shall decide whether the proposed action is suitable and whether the timeframe for implementation is appropriate. |
Art. 35 Assessment report
1 If Swissmedic approves the plans required under Article 34 paragraph 2, it shall prepare an assessment report. 2 This shall comprise the following:
|
Art. 37 Sub-contractors and subsidiaries
1 Designated bodies that delegate part of the work to sub-contractors or to a subsidiary bear full responsibility for the work carried out on their behalf by the sub-contractor or by the subsidiary. 2 They must ensure that the sub-contractor or the subsidiary meets the applicable requirements of Annex VII to EU-MDR77. 3 They must notify Swissmedic if they delegate work under the terms of paragraph 1. They must be able to demonstrate to Swissmedic that the sub-contractor or the subsidiary is capable of carrying out the tasks assigned to it. 4 Conformity assessment activities may only be delegated if the designated body has informed the legal or natural person who requested the conformity assessment accordingly. 5 The designated bodies shall make publicly available a list of their subsidiaries. 77 See the footnote to Art. 4 para. 1 let. f. |
Art. 38 Duty of cooperation and notification requirement
1 The designated bodies, including their subsidiaries and sub-contractors, are required to keep available for Swissmedic at all times all data that is necessary for assessment, designation, monitoring and re-assessment, including the documents required to assess the qualifications of sub-contractors or subsidiaries. The data must be kept up-to-date at all times. 2 The designated bodies shall notify Swissmedic within 15 days of any change that affects their ability to meet the requirements of Annex VII to EU-MDR78 or to carry out conformity assessments. 78 See the footnote to Art. 4 para. 1 let. f. |
Section 3 Suspension, Restriction or Revocation of Designation |
Art. 41 Principle
1 Designation shall be suspended, restricted or revoked if the designated body:
2 Suspensions shall be imposed for a maximum of twelve months. They may be extended by a maximum of a further twelve months. 3 If designation is suspended, restricted or revoked, the designated body must inform all affected manufacturers accordingly within ten days. |
Art. 42 Unduly issued certificates
1 In the event of its designation being restricted, suspended or revoked, the designated body shall suspend or revoke any certificates that were unduly issued. 2 If the designated body fails to fulfil this requirement, Swissmedic shall instruct it to suspend or revoke the certificates and set an appropriate deadline for doing so. |
Art. 43 Validity of certificates in the event of suspension or restriction of designation
1 If Swissmedic suspends or restricts the designation of a designated body, the certificates concerned remain valid provided Swissmedic:
2 The certificates also remain valid if Swissmedic:
3 The designated body shall notify the manufacturers concerned or the persons or entities placing the devices concerned on the market. 4 Should Swissmedic ascertain that the designated body is unable to continue to oversee existing certificates, these certificates shall retain their validity if the manufacturer of the device in question confirms to Swissmedic or, if it is domiciled in a contracting state, to the competent authority there, in writing and within three months of designation being suspended or restricted that:
79 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Art. 44 Validity of certificates in the event of designation being revoked
1 If Swissmedic revokes the designation of a designated body, the certificates affected remain valid for nine months provided:
2 Swissmedic may, within the limits of its competence, extend the provisional validity of the certificates for further periods of three months, which altogether must not exceed twelve months. |
Section 4 Monitoring and Re-assessment of Designated Bodies |
Art. 45
1 Swissmedic shall monitor the designated bodies and their subsidiaries and sub-contractors and carry out re-assessments. In the course of monitoring and re-assessing designated bodies and reviewing their assessments, Swissmedic shall take account of the requirements and procedures set out in Articles 44 and 45 EU-MDR80. 2 It shall verify whether designated bodies still satisfy the requirements of Article 36 paragraph 1 and Annex VII to EU-MDR three years after designation, and then every four years, in the course of a full re-assessment. This provision is subject to changes in assessment intervals resulting from delegated acts81 issued by the European Commission. 3 Swissmedic shall carry out an on-site audit at least once a year to ascertain whether the designated bodies and, if necessary, their subsidiaries and sub-contractors are fulfilling the requirements and obligations of Annex VII to EU-MDR. 4 For this purpose, it may at any time:
|
Chapter 6 Requirements for Economic Operators |
Section 1 Manufacturers |
Art. 46 Affixing the conformity marking and clinical evaluation
1 Manufacturers guarantee that their devices have been designed and manufactured in accordance with the requirements of this Ordinance when they place them on the market or put them into service. 2 They must affix the conformity marking to their devices. 3 They must conduct a clinical evaluation in accordance with Article 61 EU-MDR82 taking account of the amendments to this Article adopted by the European Commission by means of delegated acts83 and in accordance with Annex XIV to EU-MDR. They must update this clinical evaluation in line with the results of post-market clinical follow-up. |
Art. 47 Technical documentation
1 Manufacturers must list in the technical documentation the informationrequired in Annexes II and III to EU-MDR84, taking account of the amendments to these Annexes made by the European Commission by means of delegated acts85. 2 Manufacturers must submit either the complete technical documentation or a summary of this documentation when requested to do so by the competent authority. |
Art. 48 Document retention requirements
1 Manufacturers must ensure that the following are available to the competent authority for at least ten years after the last device covered by the declaration of conformity has been placed on the market:
2 The document retention period for implantable devices shall be at least 15 years from the date the last device was placed on the market. |
Art. 49 Person responsible for regulatory compliance
1 Manufacturers must have available within their organisation at least one person responsible for regulatory compliance who possesses the requisite expertise in the field of medical devices. 2 Proof that the person responsible for regulatory compliance possesses the requisite expertise, the responsibilities of this person, exceptions and further modalities are governed by Article 15 EU-MDR86. 3 The person responsible for regulatory compliance must have a deputy. If a number of persons are jointly responsible for regulatory compliance, their respective areas of responsibility shall be stipulated in writing. 4 The person responsible for regulatory compliance must suffer no disadvantage within the manufacturer's organisation in relation to the proper fulfilment of his or her duties, regardless of whether or not they are employees of the organisation. 86 See the footnote to Art. 4 para. 1 let. f. |
Section 2 Authorised Representative |
Art. 51 Obligations
1 Where the manufacturer of a device is not domiciled in Switzerland, the device may only be placed on the market if the manufacturer designates an authorised representative domiciled in Switzerland by means of a written mandate. 2 The authorised representative is responsible for the formal and safety-related aspects of placing the device on the market. 3 The authorised representative’s rights and obligations and the scope of its mandate are governed by Article 11 EU-MDR88. 3bis The manufacturer and authorised representative may contractually agree that instead of the authorised representative keeping available a copy of the technical documentation, the manufacturer shall, on request, submit the documentation straight to Swissmedic. The authorised representative must ensure that the documentation is submitted within seven days.89 4 Changes in authorised representative are governed by Article 12 EU-MDR. 5 Paragraphs 1–4 also apply mutatis mutandis to natural or legal persons who assemble systems and procedure packs in accordance with Article 22 paragraphs 1 and 3 EU-MDR and who are not domiciled in Switzerland.90 88 See the footnote to Art. 4 para. 1 let. f. 89 Inserted by No I of the O of 19 May 2021, in force since 26 May 2021 (AS 2021 281). 90 Inserted by No I of the O of 19 May 2021, in force since 26 May 2021 (AS 2021 281). |
Art. 52 Person responsible for regulatory compliance
1 Authorised representatives must ensure that they have permanently and continuously at their disposal at least one person who possesses the requisite expertise as regards the requirements for medical devices under this Ordinance and who is responsible for regulatory compliance. 2 In other respects, Article 49 paragraphs 2–4 shall apply mutatis mutandis. |
Section 3 Importers |
Art. 53
1 Importers may only place on the market devices that comply with this Ordinance. Before placing devices on the market, they shall verify that:
2 Importers shall indicate on the device or on its packaging or in a document accompanying the device, their name, place of business and the address at which they can be contacted. 3 Where an importer considers or has reason to believe that a device is not in conformity with the requirements of this Ordinance, it must not place the device on the market until it has been brought into conformity. 4 The further obligations of importers prior to and after placing a device on the market are governed by Articles 13 and 16 paragraphs 3 and 4 EU-MDR91. In particular, importers must comply with the following obligations:
91 See the footnote to Art. 4 para. 1 let. f. |
Section 4 Distributors |
Art. 54
1 When making a device available on the market, distributors must, in the context of their activities, act with due care in relation to the requirements applicable. Before making a device available on the market, distributors must verify that:
2 With the exception of paragraph 1 letter d, a sampling method may be used for the purposes of verification. 3 Where a distributor considers or has reason to believe that a device is not in conformity with the requirements of this Ordinance, it must not make the device available on the market until it has been brought into conformity. 4 The further obligations of distributors prior to and after making a device available on the market are governed by Articles 14 and 16 paragraphs 3 and 4 EU-MDR92. In particular, distributors must fulfil the following obligations:
92 See the footnote to Art. 4 para. 1 let. f. |
Section 5 Registration of Economic Operators93
93 Amended by No I of the O of 19 May 2021, in force since 26 May 2021 (AS 2021 281). |
Art. 55
1 Manufacturers or their authorised representatives and importers must register the information required by part A Section 1 of Annex VI to EU-MDR94 with Swissmedic within three months of placing a device on the market for the first time. 2 The economic operator in question must report any changes to the information to Swissmedic within one week. 3 Further obligations and registration modalities are governed by Articles 30 paragraph 3 and 31 EU-MDR. 4Swissmedic shall verify the information provided by the economic operators and assign them a Swiss single registration number (CHRN).95 5 Any natural or legal person placing systems and procedure packs on the market for the first time under Article 11 must register their name and the address at which they can be contacted with Swissmedic within three months of placing the system or procedure pack on the market. Where an authorised representative is required under Article 51 paragraph 5,the name and address of the authorised representative must also be registered with Swissmedic. 94 See the footnote to Art. 4 para. 1 let. f. 95 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Chapter 7 Device Surveillance |
Section 1 Post-Market Surveillance |
Art. 56 System
1 For each device, manufacturers must plan, establish, document, implement, maintain and update a post-market surveillance system in a manner that is proportionate to the risk class and appropriate for the type of device. That system shall be an integral part of the manufacturer's quality management system. 2 The system must be suited to actively and systematically gathering, recording and analysing relevant data on the quality, performance and safety of a device throughout its entire lifetime, and to drawing the necessary conclusions and to determining, implementing and monitoring any preventive and corrective actions. 3 The modalities of the system, particularly the resulting actions, updates and amendments to technical documentation, are governed by Article 83 paragraph 3 EU-MDR96. 96 See the footnote to Art. 4 para. 1 let. f. |
Art. 57 Incidents and actions
1 Should it become evident in the course of post-market surveillance that preventive and/or corrective actions are necessary, the manufacturer shall implement the appropriate measures and inform the competent authorities and, if applicable, the designated body. 2 If a manufacturer becomes aware of a serious incident in connection with a device that has been made available on the market, or takes action to prevent or minimise the risk of such an incident for medical or technical reasons (field safety corrective actions), it must report the fact in accordance with Article 66. |
Art. 59 Report
1 Manufacturers of class I devices must draw up a post-market surveillance report. 2 This report must contain:
3 The report forms part of the post-market surveillance technical documentation specified in Annex III to EU-MDR98. 4 The manufacturer must update the report when necessary and make it available to the competent authority upon request. 98 See the footnote to Art. 4 para. 1 let. f. |
Section 2 Safety Report |
Art. 60 Obligation
1 Manufacturers of class IIa, class IIb and class III devices shall prepare a safety report for each device and where relevant for each category or group of devices. 2 Manufacturers of class IIa devices shall update the safety report when necessary and at least every two years. Manufacturers of class IIb and class III devices must update this report at least annually. |
Art. 61 Content
1 The safety report must contain:
2 Throughout the lifetime of the device concerned, the safety report must set out:
3 The safety report forms part of the technical documentation specified in Annexes II and III to EU-MDR99. For custom-made devices, the report forms part of the documentation specified in Section 2 of Annex XIII to EU-MDR. 99 See the footnote to Art. 4 para. 1 let. f. |
Art. 62 Review 100
1 Manufacturers shall make their safety reports available to the designated body involved in the conformity assessment. 2 The designated body shall review the safety report for class III devices or implantable devices and record the outcome of its review with details of any action taken. 3 Manufacturers or their authorised representatives shall, upon request, make the safety report and the outcome of the designated body’s review, with details of any action taken, available to the competent authority. 100 Amended by No I of the O of 19 May 2021, in force since 26 May 2021 (AS 2021 281). |
Section 3 Summary of Safety and Clinical Performance |
Art. 63
1 For class III devices and for implantable devices, other than custom-made or investigational devices, the manufacturer must draw up a summary of safety and clinical performance. 2 The summary of safety and clinical performance shall be written in a way that is clear to the intended user and, if relevant, to the patient. 3 The minimum content of the summary is governed by Article 32, Paragraph 32 EU-MDR101. 4 The draft summary must be submitted, together with the documentation, to the designated body involved in the conformity assessment for validation by that body.102 5 The manufacturer shall publish the summary after it has been validated.103 6 The manufacturer must mention on the label or instructions for use where the summary is available. 101 See the footnote to Art. 4 para. 1 let. f. 102 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). 103 Amended by No I of the O of 19 May 2021, in force since 26 May 2021 (AS 2021 281). |
Section 4 Traceability and Recording of Device Identification |
Art. 64 Traceability 104
1 Distributors and importers shall cooperate with manufacturers or their authorised representatives to achieve an appropriate level of traceability of devices. 2 The duty of disclosure under Article 47c TPA continues for at least 10 years, or for at least 15 years for implantable devices, from the date on which the device was acquired or delivered. 104 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Art. 65 Recording the UDI
1 Economic operators and the healthcare institutions shall record and store, preferably by electronic means, the UDI of the class III implantable devices which they have supplied or with which they have been supplied.105 2 Swissmedic may extend this obligation to other devices, or categories or groups of devices. 105 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Section 5 Vigilance |
Art. 66 Reporting obligation
1 Manufacturers of devices made available in Switzerland or natural or legal persons who assemble systems or procedure packs in accordance with Article 22 paragraphs 1 and 3 EU-MDR106 must report to Swissmedic:
2 Exemptions from this reporting obligation, modalities, periodic summary reports, trend reporting and analyses of serious incidents and field safety corrective actions are governed by Articles 27 paragraph 5 and 87–89 EU-MDR. 2bis Where an authorised representative is required pursuant to Article 51, this representative is responsible for the reporting obligation in paragraph 1. Furthermore, the authorised representative must submit the trend reports pursuant to paragraph 2 on incidents in Switzerland and abroad to Swissmedic without being requested to do so. Final reports prepared in accordance with Article 89 paragraph 5 EU-MDR should be submitted to Swissmedic. The transfer of these obligations from the manufacturer or from the natural or legal person assembling systems or procedure packs under Article 22 paragraphs 1 and 3 EU-MDR to the authorised representative must be agreed in writing in the mandate.108 3 ...109 4 Any professional who becomes aware of a serious incident when using devices must report this to the supplier and Swissmedic. The report may be submitted by a professional association. The timelines for doing so are as set out in Article 87 EU-MDR. 5 Reports must be submitted to Swissmedic electronically and in machine-readable format. Swissmedic publishes information on electronic submission and the forms to be used with content specifications. 106 See the footnote to Art. 4 para. 1 let. f. 107 Amended by No I of the O of 19 May 2021, in force since 26 May 2021 (AS 2021 281). 108 Inserted by No I of the O of 19 May 2021, in force since 26 May 2021 (AS 2021 281). 109 Repealed by No I of the O of 19 May 2021, with effect from 26 May 2021 (AS 2021 281). |
Art. 67 Reporting systems in hospitals
1 Hospitals must set up an internal reporting system within the framework of an established quality management system for the purpose of reporting under Article 66 paragraph 4. 2 They must designate a suitable competent person (vigilance contact person) with a medical or technical qualification to assume responsibility for reporting to Swissmedic. They must supply this person’s contact details to Swissmedic. 3 Records and all documents created under the vigilance quality management system must be retained for at least 15 years. |
Chapter 8 Conduct in relation to Devices |
Art. 69 Advertising
1 Claims for devices must only contain statements that correspond to the product information. 2 Misleading statements, particularly concerning the intended purpose, safety and performance of a device, are prohibited. 3 Devices intended solely for use by healthcare professionals must not be advertised to the public.110 110 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Art. 70 Application 111
1 Any professional who uses a device from a foreign country directly without placing it on the market is responsible for the conformity of that device. 2 Device groups intended for use by healthcare professionals and which could harm the health of humans in the case of improper use are listed in Annex 6.112 3 The groups of devices in Annex 6 may only be used in accordance with the professional and operating requirements stated therein. 111 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). 112 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Art. 71 Maintenance
1 Any professional using devices must ensure that the devices are maintained and tested in accordance with the regulations. 2 Maintenance must be carried out in accordance with the principles of a quality management system, is to be organised appropriately, and must be guided in particular by:
3 For devices with a measurement function, test procedures may be required in accordance with the Measuring Instruments Ordinance of 15 February 2006113. 4 Swissmedic may issue and publish requirements for maintenance measures. These requirements shall be deemed to constitute the current scientific and technological standards. |
Art. 72 Reprocessing
1 Any person using in a professional capacity a device intended for repeated use must ensure on each occasion and prior to use that its functionality has been tested and that it has been reprocessed in accordance with current scientific and technological standards and taking account of the instructions of the manufacturer and the requirements of hygiene. 2 Reprocessing must employ suitable procedures that have been validated in accordance with current scientific and technological standards and whose efficacy has been demonstrated and can be reliably traced and reproduced within a quality management system. 3 Any natural or legal person who reprocesses devices for third parties must:
4 The declaration required under paragraph 3 letter a must identify the device and state the name and address of the establishment that reprocessed it. |
Art. 74 Cyber security
1 Healthcare institutions must put in place all technical and organisational resources required by the state of the art to ensure that network-compatible devices are protected against electronic attack and unauthorised access. 2 Hospitals must identify, evaluate and document the measures taken under paragraph 1 in accordance with the principles of a risk management system. This system forms an integral part of the hospitals’ quality management system. |
Chapter 9 Market Surveillance |
Art. 75 Principle
1 Inspections under the auspices of market surveillance shall cover devices made available on the market, conformity assessment procedures, device surveillance, device handling and economic operators’ fulfilment of their obligations. They shall also cover devices made available in contracting states by natural or legal persons domiciled in Switzerland, the conformity assessment procedures and surveillance activities for such devices and the natural or legal persons’ fulfilment of their obligations. 2 The market surveillance activities undertaken by Swissmedic and the Cantons are governed by Article 66 TPA and Articles 93–95, 97 and 98 EU MDR115. Articles 97 paragraph 3 and 98 paragraphs 3 and 4 EU MDR are excluded.116 3 The Cantons shall draw up annual plans for their market surveillance activities under paragraph 2. They shall provide Swissmedic with an annual summary of the results of their surveillance activities. Swissmedic may determine both the content of the summary and the form in which it is made available. 4 In case of an actual necessity for the protection of public health, Swissmedic shall decree the measures under Article 66 TPA in a general ruling. 115 See footnote to Art. 4 para. 1 let. f. 116 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Art. 75a Common activities and use of information 117
117 Inserted by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Art. 75b Additional measures 118
118 Inserted by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Art. 76 Responsibilities
1 Swissmedic is responsible for monitoring:
2 Certain aspects of the monitoring activities set out in paragraph 1 remain the responsibility of other federal offices or institutions. 3 The Cantons are responsible for monitoring:
|
Art. 77 Powers
1 For the purposes of verifying conformity, the authorities responsible for monitoring under Article 76 may, without providing compensation:
2 If a manufacturer or a natural or legal person who assembles systems or procedure packs under Article 22 paragraphs 1 and 3 EU-MDR119 fails to fulfil their obligations under Article 66, Swissmedic may impose appropriate measures to protect health, up to and including prohibiting the making available on the market or the putting into service of the devices in question.120 119 See the footnote to Art. 4 para. 1 let. f. 120 Inserted by No I of the O of 19 May 2021, in force since 26 May 2021 (AS 2021 281). |
Art. 78 Duty to cooperate and provide information
1 Economic operators that place a device on the market in Switzerland or in a contracting state, and economic operators, professionals and healthcare institutions that make a device available or put it into service in Switzerland or a contracting state have a duty to cooperate on matters of enforcement. In particular, they must provide, free of charge, all necessary information and all necessary proof and documentation to the enforcement bodies. 2The providers of information society services also have a duty to cooperate on matters of enforcement. In particular, they should inform the competent authorities about suspected illegal activities by, or information from, users of their service and, upon request, provide information that enables the users of their service with whom they have concluded agreements about storage to be identified.121 121 Inserted by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Chapter 10 Data Processing |
Section 2 Medical Devices Information System |
Art. 84 Data protection and security
1 Swissmedic shall draw up processing rules in accordance with Article 6 of the Data Protection Ordinance of 31 August 2022122 (DPO).123 2 Articles 1–4 und 6 DPO apply to data security.124 3 Data processing must be automatically logged. 123 Amended by Annex 2 No II 101 of the Data Protection Ordinance of 31 Aug. 2022, in force since 1 Sept. 2023 (AS 2022 568). 124 Amended by Annex 2 No II 101 of the Data Protection Ordinance of 31 Aug. 2022, in force since 1 Sept. 2023 (AS 2022 568). |
Art. 85 Content of the medical devices information system
1 This medical devices information system contains the following data:
2 The medical devices information system only contains personal data where such data are required to record and process information. |
Art. 86 Data exchange with other information systems 125
The medical devices information system may draw the data referred to in Article 85 from theEuropean Database on Medical Devices and from cantonal electronic systems; it may also submit data to Eudamed and cantonal systems. 125 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Art. 88 Data subjects’ rights and data rectification
1 Data subjects’ rights, particularly the right to information, rectification and deletion of data, are governed by data protection legislation. 2 Swissmedic shall ensure that data that are incorrect or have been processed unlawfully are corrected in or deleted from the medical devices information system. Correction and deletion shall take place as quickly as possible, but no later than 60 days of the data subject making their request. |
Art. 89 Data archiving
The personal data specified in Article 85 paragraph 2 must be stored in such a way that it is only possible to identify data subjects for a period of ten years after the last device covered by the declaration of conformity was placed on the market. For implantable devices, this period is extended to 15 years. |
Art. 90 Publication of data
Swissmedic may publish the following in the medical devices information system:
126 See the footnote to Art. 4 para. 1 let. f. |
Art. 92 Applicability of the Data Protection Act 127
All data processing activities carried out in the medical devices information system must comply with the Data Protection Act of 25 September 2020128. 127 Amended by Annex 2 No II 101 of the Data Protection Ordinance of 31 Aug. 2022, in force since 1 Sept. 2023 (AS 2022 568). |
Chapter 11 Final Provisions |
Section 1 Enforcement |
Art. 93 Amendment of Annexes
1 The FDHA may amend Annexes 1–3, 5 and 6 to this Ordinance in line with international and technical progress. 2 Where amendments may pose technical barriers to trade, it shall effect them by mutual agreement with the Federal Department of Economic Affairs, Education and Research. |
Art. 94 Information on directly applicable legal acts of the European Commission
Swissmedic shall provide on its website information on legal acts of the European Commission that, in accordance with this Ordinance, are directly applicable in Switzerland in the version binding upon the Member States of the EU and as listed in Annex 4. |
Art. 95 Harmonisation of enforcement
1 When implementing this Ordinance, Swissmedic shall respect the implementing acts adopted by the European Commission on the basis of EU-MDR129. 2 Regulations (EU) No 207/2012130 and No 722/2012131 shall remain in force until such time as they are repealed in the EU by implementing acts adopted by the European Commission on the basis of EU-MDR. 129 See the footnote to Art. 4 para. 1 let. f. 130 Commission Regulation (EU) No 207/2012 of 9 March 2012 on electronic instructions for use of medical devices, Amended by OJ L 72 of 10.3.2012, p. 28. 131 Commission Regulation (EU) No 722/2012 of 8 August 2012 concerning particular requirements as regards the requirements laid down in Council Directives 90/385/EEC and 93/42/EEC with respect to active implantable medical devices and medical devices manufactured utilising tissues of animal origin, Amended by OJ L 212 of 9.8.2012, p. 3. |
Art. 96 Cooperation with the European Commission and authorities of the contracting states
1 Where provided for by international agreements, Swissmedic, the designated bodies and economic operators as defined in Art. 47 TPA shall cooperate with the European Commission and the authorities of the contracting states.132 2 Swissmedic may appoint experts who are qualified to evaluate conformity assessment bodies in the field of medical devices. 3 Swissmedic may appoint expertsto participate in expert groups of the European Commission and the authorities of the contracting states. 132 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Art. 97 Collaboration with the customs authorities
1 The customs authorities provide Swissmedic with information on the import, export and transit of devices. 2 Swissmedic may mandate the customs authorities to detain devices for further inspection and to obtain samples. 3It may provide the customs authorities with information about ongoing or concluded administrative or criminal proceedings and sanctions in connection with market surveillance.133 133 Inserted by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
Art. 98 Expert laboratories in Switzerland
1 Laboratories that wish to be designated an expert laboratory by the European Commission in accordance with Article 106 paragraph 7 EU-MDR134 may apply to Swissmedic for designation. 2 They must demonstrate to Swissmedic in particular that they:
3 They must operate in one of the following fields:
4 If the requirements are met, Swissmedic shall propose to the European Commission that the laboratory be designated an expert laboratory. |
Section 2 Repeal of other Legislation and Transitional Provisions |
Art. 99 Repeal of other legislation
The following ordinances are repealed:
136 [AS 2001 3487;20044037no I 5, II para. 2; 20084377Annex 5 no 2; 20101215, 2749no I 7; 2015 999; 2017 5935; 2019 999Art. 28 para. 2; 2020 2975] |
Art. 100 Validity of certificates issued under the old legislation
1 Certificates issued under the old legislation prior to 25 May 2017 shall retain their validity until the expiry date stated therein, but no longer than 26 May 2022. 2 Certificates issued under the old legislationsince 25 May 2017 shall retain their validity until the expiry date stipulated therein, but no longer than 26 May 2024. |
Art. 101 Placing on the market of products that comply with the old legislation
1 Provided they continue as of 26 May 2021 to comply with the old legislation and have not undergone any significant changes in their design or intended purpose, the following medical devices may be placed on the market or put into service until 26 May 2024:
2 The post-market surveillance and market surveillance of these devices, vigilance, and registration of economic operators and of the devices themselves are subject to the provisions of this Ordinance. 3 Devices legally placed on the market prior to 26 May 2021 under the old legislation and devices placed on the market from 26 May 2021 under paragraph 1 may continue to be made available on the market or put into service until 26 May 2025. The above is subject to Article 103. |
Art. 103 Devices incorporating devitalised tissues or cells of human origin
1 Devices incorporating devitalised tissues or cells of human origin or their derivatives as specified in Article 1 paragraph 3 letters c no. 2 and d that were lawfully placed on the market or put into service prior to 26 May 2021 may continue to be made available on the market or put into service until 26 May 2025. Art. 101 paragraph 2 applies mutatis mutandis. 2 Until such time as a corresponding special Ordinance is issued, devices covered by Article 2a paragraph 2 TPA are subject to the Medical Devices Ordinance of 17 October 2001139. Devices covered by Article 2a paragraph 2 TPA that were lawfully placed on the market prior to 26 May 2021 may continue to be made available on the market or put into service until such time as a corresponding special Ordinance is issued. 139 See the footnote to Art. 99 no 1. |
Art. 104 Affixing the UDI
The UDI required by Article 17 paragraph 2 must be affixed:
|
Art. 104a Designation of an authorised representative 140
1 If the manufacturer is domiciled in an EU or EEA state or if the manufacturer has designated an authorised representative domiciled in an EU or EEA state, that manufacturer must, for all devices placed on the market after 26 May 2021, designate an authorised representative in accordance with Article 51 paragraph 1 within the following time periods:
2 For systems and procedure packs, an authorised representative in accordance with Article 51 paragraph 5 must be appointed by 31 July 2022. 140 Inserted by No I of the O of 19 May 2021, in force since 26 May 2021 (AS 2021 281). |
Art. 104abis Placing information about the authorised representative 141
For class I devices and for systems and procedure packs that are placed on the market in accordance with the new legislation, the information specified in Annex 1 Chapter III Section 23.2 letter d EU-MDR142 about the authorised representative in accordance with Article 51 paragraph 1 or 5 of this Ordinance may be included in a document accompanying the device until 31 July 2023. 141 Inserted by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). 142 See the footnote to Art. 4 para. 1 let. f. |
Art. 104b143
143 Inserted by No I of the O of 19 May 2021 (AS 2021 281). Repealed by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, with effect from 26 Mai 2022 (AS 2022 291). |
Art. 105144
144 Repealed by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, with effect from 26 May 2022 (AS 2022 291). |
Art. 107 Conformity assessment bodies
1 Conformity assessment body designations issued under Section 4 of the Medical Devices Ordinance of 17 October 2001145 shall become void. 2 The conformity assessment body that issued the certificates under the old legislation shall remain responsible for the appropriate surveillance of these devices. It shall be subject to supervision by Swissmedic. 3 Conformity assessment body designations issued under Section 4a of the Medical Devices Ordinance of 17 October 2001 shall retain their validity. 4 ...146 145 See the footnote to Art. 99 no 1. 146 Repealed by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, with effect from 26 May 2022 (AS 2022 291). |
Art. 108 Notification of devices, systems and procedure packs 147
1 Until Article 17 paragraph 5 enters into force, the following notification obligations shall continue to apply to natural or legal persons domiciled in Switzerland:
2 ...150 147 Amended by No I of the O of 19 May 2021, in force since 26 May 2021 (AS 2021 281). 148 See the footnote to Art. 4 para. 1 let. f. 149 See the footnote to Art. 99 no 1. 150 Enters into force at a later date (Art. 110 para. 2). |
Art. 109151
151 Repealed by No I of the O of 19 May 2021, with effect from 26 May 2021 (AS 2021 281). |
Art. 110 Entry into force
1 Subject to the exceptions in paragraph 2, this Ordinance enters into force on 26 May 2021. 2 Articles 17 paragraph 5 and 108 paragraph 2 enter into force at a later date.152 152 Amended by No I of the O of 19 May 2021, in force since 26 May 2021 (AS 2021 281). |
Annex 1 |
(Art. 1 para. 1 letter b) |
Annex 2 153
153 Amended by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
(Art. 5 para. 1) |
Equivalent terms |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
The terms listed below and used in EU MDR154 and this Ordinance are equivalent as follows:
154 See the footnote to Art. 4 para. 1 let. f. |
|
Annex 3 155
155 Revised by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices (AS 2022 291) and Annex 2 No II 101 of the Data Protection Ordinance of 31 Aug. 2022, in force since 1 Sept. 2023 (AS 2022 568). |
(Art. 5 para. 2) |
Applicable law |
1 EU law |
Where this Ordinance makes reference to provisions of EU MDR 156 that in turn make reference to the following EU law, the version below remains applicable:
156 See the footnote to Art. 4 para. 1 let. f. |
2 Swiss law |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Where this Ordinance makes reference to provisions of EU-MDR157 that in turn make reference to EU law, the Swiss law below is applicable in place of the EU law:
|
Annex 4 172
172 Revised by the correction of 4 May 2021 (AS 2021 260). |
(Art. 4 para. 2, 17 para. 4, 19 para. 1 let. c, 20 para. 1, 23, 25 para. 3, 29 para. 2, 45 para. 2, 46 para. 3, 47 para. 1 and 98 para. 2 let. b) |
Delegated acts of the European Commission based on EU-MDR 173
173 See the footnote to Art. 4 para. 1 let. f. |
||||||||||||||||||||||||
For the purposes of implementing this Ordinance, the legal acts adopted on the basis of the provisions of EU-MDR set out below shall apply in Switzerland in the binding version applicable to the EU Member States:
|
Annex 5 |
(Art. 13 para. 1) |
Conformity marking |
The conformity marking is as follows: Where a designated body has to be involved, its identification number is to be placed beside its conformity marking. |
Annex 6 174
174 Revised by Annex 5 No 1 of the O of 4 May 2022 on In Vitro Diagnostic Medical Devices, in force since 26 May 2022 (AS 2022 291). |
(Art. 70, paras 2 and 3) |
Device Groups Intended for Use by Healthcare Professionals, Qualifications and Circumstances of Use |